Volume 15 , Issue 4 , September 2013 , Pages 22-28
1 School of Science Education, University of Sulaimani
Investigation in doses delivered to patient during an X-ray imaging is carried out. From the data obtained, the Entrance Surface Dose (ESD) is calculated and compared to the Reference Dose Levels of other countries (UK, Australia and the US). The results show 93% of chest examination was cases of overdose, while some other examinations had 8.5% overdose by more than 3 times. The results suggests that during X-ray imaging patients receive doses more than the reference level at the Sulaimani Hospital.