Investigation into Radiation Doses to Patient During Diagnostic Imaging

Volume 15 , Issue 4 , September 2013 , Pages 22-28

Authors

Dastan Khalid 1

1 School of Science Education, University of Sulaimani

DOI logo 10.17656/jzs.10270

Keywords

Abstract


Investigation in doses delivered to patient during an X-ray imaging is carried out. From the data obtained, the Entrance Surface Dose (ESD) is calculated and compared to the Reference Dose Levels of other countries (UK, Australia and the US). The results show 93% of chest examination was cases of overdose, while some other examinations had 8.5% overdose by more than 3 times. The results suggests that during X-ray imaging patients receive doses more than the reference level at the Sulaimani Hospital. 